해외 우수한 분석, 측정, 계측기기 제조회사의 독점 대리점 !
Model | MP25 |
---|---|
Spectral Response | 2.0 - 2.8 μm |
Temperature Range | 75 –500 °C |
100 – 700 °C | |
160 – 1200 °C | |
200 – 1300 °C |
Pyrometer | Specification |
---|---|
Measurement uncertainty | < 400°C: 2°C; > 400°C: 0,3% of measured value in °C + 1 K (TA = 23°C, ε = 1, t90 = 1 s) |
Repeatability | 0.1% of measured value in °C + 1 K (TA = 23°C, ε = 1, t90 = 1 s) |
Response time t90 | 3 ms with dynamic adaptation at low signal levels, adjustable up to 10 s |
Emissivity adjustable range | 0.05 - 1.00 |
Analog output signal | 0 or 4 ~ 20 mA selectable, 500 Ω max. load |
Digital Interface | Either RS232 or RS485 max. 57.6 kBd (standard supply), Profibus (optional) |
Temperature resolution | Analog < 0.1% of adjusted temperature range, digital 0.1°C |
Ambient temperature | Operation 0 - 53°C, storage -20 - 60°C |
Rel. Humidity | No condensing conditions |
Power Supply | 24 V AC/DC (15 - 30 V AC/DC), AC: 48 - 62 Hz, max. 7 VA |
Isolation | Power supply, analog and digital output are galvanically isolated against each other and against housing |
Housing and protection class | Aluminum, IP65 per DIN 40 050 |
Weight | 600 g |
CE Label | According to EU directives for electromagnetic immunity |
Laser targeting light | (optional) 650 nm, < 1 mW, class II per IEC 60825-1-3-4 |
Focusable Lens | |||||
---|---|---|---|---|---|
Lens | Distance | Spot Size Diameter | |||
75-550°C | 100-700°C | 160-1200°C | 200-1300°C | ||
OM25-A0 | 100 mm | 0.85 mm | 0.6 mm | 0.4 mm | 0.25 mm |
120 mm | 1.3 mm | 0.85 mm | 0.5 mm | 0.35 mm | |
133 mm | 1.6 mm | 1.0 mm | 0.6 mm | 0.4 mm | |
OM25-B0 | 190 mm | 1.5 mm | 1.0 mm | 0.7 mm | 0.35 mm |
260 mm | 2.4 mm | 1.6 mm | 1.1 mm | 0.6 mm | |
340 mm | 3.4 mm | 2.3 mm | 1.6 mm | 0.8 mm | |
OM25-C0 | 350 mm | 2.7 mm | 1.85 mm | 1.25 mm | 0.65 mm |
1200 mm | 11.5 mm | 7.8 mm | 5.2 mm | 2.6 mm | |
2600 mm | 26.0 mm | 17.5 mm | 11.6 mm | 5.8 mm |